BGO setup
NOTE: to access the FERA pedestals in crate #2 two things
must be fulfilled:
- the MSU data acquisition must be stopped. Type 'end'
in the FE control window.
- the MSU system may leave the CAMAC crate in a mode where CNAF
commands cannot be issued. Use the command 'reset_camac lta8xxx:' to
reset the crate (see the
MSU online documentation for the latport numbers). The command is
automatically issued in e.g., the auto_cal.com and tcheck.com files, so
you may not have to issue it by hand. Be sure to issue it before you
use bgo_setup to save your parameters and check that the save was ok
using the list_save utility.
- BGO energy calibration
- BGO time calibration
- BGO walk and threshold calibration
- MLU setup
- To set the PLD (programmable delay module) in crate 1 do:
- stop the MSU data acquisition (end)
- stop the DAPHNE acquisition (stp)
- in the [bgo.daq1] directory type @PLD
- You will have to change the latport in the
first line of the command file if you are
using a different acquisition station than
the script expects(see the
MSU online documentation for the latport
numbers).
-
- To set the BGO DTG (dead time generators) in crate 4 do:
- stop the MSU data acquisition (end)
- stop the DAPHNE acquisition (stp)
- in the [bgo.daq1] directory type @SET_DTG
- You will have to change the latport in the
first line of the command file if you are
using a different acquisition station than
the script expects(see the
MSU online documentation for the latport
numbers).
-