BGO setup


1) collect time spectra after they have been aligned. Write them out by "WR_THEX".

2) Get into GAGA. Gaga should be defined on ATLA but otherwise "@[replay.bgo.hans.analyze]gaga" will do it. Use DRE to get the first spectrum in (actually to define the file name for the following command files). Now use "CF SUM_THEX" and "CF SUM_THEX_2" in the [.current_bgo_cal] directory to actually add up the spectra. Now make a hard copy and mark the gate limits.

3) to set the MLU do: type BGO_SETUP and select MLU. An example of the setup from here is:


        LMEM 1           . specifies which MLU (allways one)
        N                . not discriminator mode
        5                . we ask for 5 windows
        404,580,2048     . background marked 2048
        684,902,64       . three isomer windows
        974,1184,64      . _ marked 64
        1268,1442,64     . _
        1443,1576,32769  . prompt window marked 32769

     note: it takes a while to load the memory of the MLU

there are 82.41 nsec between beam pulses.
web master: [email protected]